top of page

National Diagnostic Reference Levels (NDRLs)

GUIDANCE:

National Diagnostic Reference Levels (NDRLs)

Qatar 2018

Date of Issue

Date of Review

: 28 October 2018

: 25 October 2021

The following tables list the National Diagnostic Reference Levels (NDRLs) for the State of Qatar. These NDRL values are selected from several national and regional RDLs and approved by Qatar Medical Physicists Society (QaMPS) in 2018. They should be considered by employers when setting their local DRLs as required by the Executive Regulations (Dec. 4 2003) of the Radiation Protection Law 31 (2002).

​

                                              Signed:

​

                                                                                                             

​

Dr. Huda AlNaemi

                                 

QaMPS President

​

CONTENTS:

​

1.0    Qatar DRLs 2018 for Computed Tomography

         1.1  Qatar DRLs 2018 for for Adult Computed Tomography

          1.2  Qatar DRLs 2018 for Pediatric Computed Tomography

​

2.0    Qatar DRLs 2018 for General Radiography

           2.1  Qatar DRLs 2018 for Adult General Radiography

         2.2  Qatar DRLs 2018 for Pediatric General Radiography

         2.3  Qatar DRLs 2018 for Mammography

​

3.0    Qatar DRLs 2018 for Mammography

​

4.0    Qatar DRLs 2018 for Fluoroscopy

​

5.0    Qatar DRLs 2018 for Nuclear Medicine

​

6.0    Qatar DRLs 2018 for 8F- fluorodeoxyglucose Positron Emission Tomography (18F-FDG PET)

​

Appendices

References

1.0  Qatar DRLs 2018 for Computed Tomography

(1)

1.1    Qatar DRLs 2018 for Adult Computed Tomography

1.1 Qatar DRLs 2018 for Adult Computed T

1.2    Qatar DRLs 2018 for Pediatric Computed Tomography

(3)

1.2 Qatar DRLs 2018 for Pediatric Comput

2.0  Qatar DRLs 2018 for General Radiography

2.1    Qatar DRLs 2018 for Adult General Radiography

(4)

2.1 Qatar DRLs 2018 for Adult General Ra

2.2   Qatar DRLs 2018 for Pediatric General Radiography

(3)

2.2 Qatar DRLs 2018 for Pediatric Genera

(6)

3.0  Qatar DRLs 2018 for Mammography

Mean glandular dose: 2.4 mGy

4.0  Qatar DRLs 2018 for Fluoroscopically Guided Interventional Procedures

(6)

Fluoroscopic radiation dose rate: 20 mGy/min

5.0  Qatar DRLs 2018 for Nuclear Medicine

(7)

5. Qatar DRLs 2018 for Nuclear Medicine

6.0 Qatar DRLs 2018 for 18F-fluorodeoxyglucose Positron Emission Tomography (18F-FDG PET) 

(7)

6. Qatar DRLs 2018 for 18F-fluorodeoxygl

*Kg is the weight of the patient in Kg

Appendix

Approximate equivalence of weight and age groups for the purpose of pediatric DRLs

(3)

Appendix..JPG

References

​

1.   Unless otherwise indicated, DRLs for Adult CT examinations from Australia (July 2018) as published

      by ARPANSA online [Accessed September 2018]

​

2.   DRL for CTA from United Kingdom PHE-CRCE-013: Doses from computed tomography (CT)

      examinations in the UK (2011 Review) and published online [Accessed September 2018]

​

3.   Radiation Protection Report No. 185, European Guidelines on Diagnostic Reference levels for

      Paediatric Imaging (2018)

​

4.   Unless otherwise indicated, DRLs for adult general X-ray examinations from United Kingdom

      HPA-CRCE-034: Doses to patients from radiographic and fluoroscopic x-ray imaging procedures

      in the UK (2010 review) and published online [Accessed September 2018]

​

5.   Federal Office for Radiation Protection in German, The updated diagnostic reference values for       

      diagnostic and interventional X-ray applications, vom 22. June 2016

​

6.   Japan Association on Radiological Protection in Medicine, Diagnostic Reference Levels Based on

      Latest Surveys in Japan, 2015

​

7.   Australian Radiation Protection and Nuclear Safety Agency, Nuclear Medicine Diagnostic Reference

      Levels (DRLs), 2017

​

home icon.png

Home   /   About   /   NDRLs

About QaMPS
     Constitution & By-Laws
     Qatar NDRLs
     Publications

Useful Links

     MEFOMP

     IOMP

     IAEA 

     IMPCB 

​

NEW

Nuklearmedizin.jpg

To be posted soon.

.
bottom of page